Functional Characteristics |
|||
Temperature Control |
Humidity Control |
||
Air Control Range |
Normal : 23.0 ~ 37.0°C |
Control Range |
30 ~ 95% |
Override : 37.1 ~ 39.0°C |
Measurement Range |
15~99% ±5% RH(non-condesing) |
|
Air Measurement range |
20.0 ~ 45.0°C ±0.5°C |
Control System |
Steam(Boiling at 100°C) |
Skin Control Range |
Normal : 35.0 ~ 37.5°C |
Water Tank |
1,000ml (Autoclavable) |
Override : 37.6 ~ 39.0°C |
Display |
|
|
Skin Measurement range |
25.0 ~ 45.0°C ±0.3°C |
Type |
7″ TFT Color LCD |
Function |
Hood |
|
|
Microprocessor Controlled & Double Safety Feature |
Mattress tilt |
up to 12° |
|
Trend |
9~72hours (Weighing Scale 9days) |
Noise Level |
< 45dB |
Alarm(Visual & Sound) |
27 Events Alarms |
Air Filter Particle Size |
0.3micron |
Warm-up Time |
Silent Mode : 60min, <45dB |
Air Filter Efficiency |
99.8% |
Normal Mode : 45min, <55dB |
Double Wall |
||
Fast Mode : 30min, <66dB |
Air Curtain |
||
Alarm Sounds for 30min when Power Outage |
X-ray Tray |
||
Multi-language |
Masimo SpO2 with CCD & External Monitor (Option) |
||
Quick Guide |
Display |
7″ TFT Color LCD |
|
Power On Self Test |
SpO2 Range |
1~ 100% |
|
O2 Measurement (Option) |
SpO2 Accuracy |
70 ~ 100% ±3digits |
|
Measurement Range |
18 ~ 100% ±5% |
0 ~ 69% : Unspecified |
|
O2 Servo Control (Option) |
Pulse Rate Range |
25 ~ 240bpm ±3digits |
|
Control Range |
21 ~ 65% ±5% |
CCD Camera |
510 x 492 pixels |
Weighing Scale (Option) |
Video Recording |
||
Measurement Range |
0 ~ 10.0Kg ±10g |
||
Shelf (Option) |
Basket (Option) |
||
Weight Limit |
Approx. 3Kg |
Weight Limit |
Approx. 10Kg |
IV Plate (Option) |
PC Interface |
||
Weight Limit |
Approx. 11Kg |
RS232C |
|
Power |
|||
Input |
AC 110/220, 50/60Hz |
Consumption |
950VA |
Output |
AC 110/220, 50/60Hz, Max.400VA |
Rechargeable Battery |
Li-ion, 3.7V, 2,200mA |
Standard Configuration |
|||
Skin Temperature Sensor |
2ea |
Air Filter |
4ea |
Power Cord |
1ea |
Operation manual |
1ea |
Options |
|||
Lifting Stand |
Ringer Pole |
||
IV Plate |
Weighing Scale |
||
Basket |
Shelf |
||
O2 Monitoring |
O2 Servo Control (O2 Monitoring Included) |
||
Masimo SpO2 with CCD & External Monitor |
|||
Warranty |
|||
Main Unit |
2years |
||
Physical Characteristics |
|||
Dimension |
Weight |
||
Standard |
1,024(W) x 690(D) x 1,354(H)mm |
Standard / Full Option |
99.3Kg / 117.5Kg |
Stand Fixed / Lifting |
645(H) mm / 651 ~ 851(H)mm |
||
Mattress |
730(W) x 380(D) x 27(H)mm |
||
Packing |
1,150(W) x 800(D) x 1,530(H)mm |
Packing |
126Kg(Full Option) |
Environmental Conditions |
|||
Operating Temperature |
20 to 30°C (68 ~ 86°F) |
Storage Temperature |
–20 to 60°C (–4 ~ 140°F) |
Operating Humidity |
0 ~ 95% non-condensing |
Storage Humidity |
0 ~ 95% non-condensing |
Certificates |
|||
Certificates |
KFDA, CE, FDA |
Classifications |
Class II, Type BF |
Standard |
EN60601-1, EN60601-1-2 |
||
EN 60601-2-19 |